MODEL SPECIFICATIONS:
Single-channel Adjustable Volume Pipettes | |||||||
Model No | Volume Range | Increment | Test Volume (μl)
|
Maximum Permissible
Systematic error |
Maximum Permissible
Random error |
||
% | μl | % | μl | ||||
MM2.5 |
0.1-2.5μl |
0.05μl |
2.5 | 2.50% | 0.0625 | 2.00% | 0.05 |
1.25 | 3.00% | 0.0375 | 3.00% | 0.0375 | |||
0.25 | 12.00% | 0.03 | 6.00% | 0.015 | |||
MM10 |
0.5-10μl |
0.1μl |
10 | 1.00% | 0.1 | 0.80% | 0.08 |
5 | 1.50% | 0.075 | 1.50% | 0.075 | |||
1 | 2.50% | 0.025 | 1.50% | 0.015 | |||
MM20 |
2-20μl |
0.5μl |
20 | 0.90% | 0.18 | 0.40% | 0.08 |
10 | 1.20% | 0.12 | 1.00% | 0.1 | |||
2 | 3.00% | 0.06 | 2.00% | 0.04 | |||
MM50 |
5-50μl |
0.5μl |
50 | 0.60% | 0.3 | 0.30% | 0.15 |
25 | 0.90% | 0.225 | 0.60% | 0.15 | |||
5 | 2.00% | 0.1 | 2.00% | 0.1 | |||
MM100 |
10-100μl |
1μl |
100 | 0.80% | 0.8 | 0.15% | 0.15 |
50 | 1.00% | 0.5 | 0.40% | 0.2 | |||
10 | 3.00% | 0.3 | 1.50% | 0.15 | |||
MM20/200 |
20-200μl |
1μl |
200 | 0.60% | 1.2 | 0.15% | 0.3 |
100 | 0.80% | 0.8 | 0.30% | 0.3 | |||
20 | 3.00% | 0.6 | 1.00% | 0.2 | |||
MM50/200 | 50-200μl |
1μl |
200 | 0.60% | 1.2 | 0.15% | 0.3 |
100 | 0.80% | 0.8 | 0.30% | 0.3 | |||
50 | 1.00% | 0.5 | 0.40% | 0.2 | |||
MM100/1000 |
100-1000μl |
5μl |
1000 | 0.60% | 6 | 0.20% | 2 |
500 | 0.70% | 3.5 | 0.25% | 1.25 | |||
100 | 2.00% | 2 | 0.70% | 0.7 | |||
MM200/1000 |
200-1000μl |
5μl |
1000 | 0.60% | 6 | 0.20% | 2 |
500 | 0.70% | 3.5 | 0.25% | 1.25 | |||
200 | 0.90% | 1.8 | 0.30% | 0.6 | |||
MM5000 |
1000-5000μl |
50μl |
5000 | 0.50% | 25 | 0.15% | 7.5 |
2500 | 0.60% | 15 | 0.30% | 7.5 | |||
1000 | 0.70% | 7 | 0.30% | 3 | |||
MM10ml |
2-10ml |
0.1ml |
10ml | 0.60% | 60 | 0.20% | 20 |
5ml | 1.20% | 60 | 0.30% | 15 | |||
2ml | 3.00% | 60 | 0.60% | 12 |
8-channel Adjustable Volume Pipettes | |||||||
Model No | Volume Range | Increment | Test Volume (μl)
|
Maximum Permissible
Systematic error |
Maximum Permissible
Random error |
||
% | μl | % | μl | ||||
MM10/8 |
0.5-10μl |
0.1μl |
10 | 1.50% | 0.15 | 1.50% | 0.15 |
5 | 2.50% | 0.125 | 2.50% | 0.125 | |||
1 | 4.00% | 0.04 | 4.00% | 0.04 | |||
MM50/8 |
5-50μl |
0.5μl |
50 | 1.00% | 0.5 | 0.50% | 0.25 |
25 | 1.50% | 0.375 | 1.00% | 0.25 | |||
5 | 3.00% | 0.15 | 2.00% | 0.1 | |||
MM300/8 | 50-300μl |
5μl |
300 | 0.70% | 2.1 | 0.25% | 0.75 |
150 | 1.00% | 1.5 | 0.50% | 0.75 | |||
50 | 1.50% | 0.75 | 0.80% | 0.4 |
12-channel Adjustable Volume Pipettes | |||||||
Model No | Volume Range | Increment | Test Volume (μl)
|
Maximum Permissible
Systematic error |
Maximum Permissible
Random error |
||
% | μl | % | μl | ||||
MM10/12 |
0.5-10μl |
0.1μl |
10 | 1.50% | 0.15 | 1.50% | 0.15 |
5 | 2.50% | 0.125 | 2.50% | 0.125 | |||
1 | 4.00% | 0.04 | 4.00% | 0.04 | |||
MM50/12 |
5-50μl |
0.5μl |
50 | 1.00% | 0.5 | 0.50% | 0.25 |
25 | 1.50% | 0.375 | 1.00% | 0.25 | |||
5 | 3.00% | 0.15 | 2.00% | 0.1 | |||
MM300/12 |
50-300μl |
5μl |
300 | 0.70% | 2.1 | 0.25% | 0.75 |
150 | 1.00% | 1.5 | 0.50% | 0.75 | |||
50 | 1.50% | 0.75 | 0.80% | 0.4 |
Fixed Volume Pipettes | |||||||
Model No | Volume Range | Increment | Test Volume (μl)
|
Maximum Permissible
Systematic error (Inaccuracy) |
Maximum Permissible
Random error (Inaccuracy) |
||
% | μl | % | μl | ||||
MM5F | 5μl | – | 5μl | 1.3% | 0.065 | 1.2% | 0.06 |
MM10F | 10μl | – | 10μl | 0.8% | 0.08 | 0.8% | 0.08 |
MM20F | 20μl | – | 20μl | 0.6% | 0.12 | 0.5% | 0.1 |
MM25F | 25μl | – | 25μl | 0.5% | 0.125 | 0.3% | 0.075 |
MM50F | 50μl | – | 50μl | 0.5% | 0.25 | 0.3% | 0.15 |
MM100F | 100μl | – | 100μl | 0.5% | 0.5 | 0.3% | 0.3 |
MM200F | 200μl | – | 200μl | 0.4% | 0.8 | 0.2% | 0.4 |
MM250F | 250μl | – | 250μl | 0.4% | 1.0 | 0.2% | 0.5 |
MM500F | 500μl | – | 500μl | 0.3% | 1.5 | 0.2% | 1.0 |
MM1000F | 1000μl | – | 1000μl | 0.3% | 3.0 | 0.2% | 2.0 |
MM2000F | 2000μl | – | 2000μl | 0.3% | 6.0 | 0.15% | 3.0 |
MM5000F | 5000μl | – | 5000μl | 0.3% | 15 | 0.15% | 7.5 |