MODEL SPECIFICATIONS:
Single-channel Adjustable Volume Pipettes | |||||||
Model No | Volume Range | Increment | Test Volume (μl)
|
Maximum Permissible
Systematic error |
Maximum Permissible
Random error |
||
% | μl | % | μl | ||||
MP2.5A |
0.1-2.5μl |
0.05μl |
2.5 | 2.50% | 0.0625 | 2.00% | 0.05 |
1.25 | 3.00% | 0.0375 | 3.00% | 0.0375 | |||
0.25 | 12.00% | 0.03 | 6.00% | 0.015 | |||
MP10A |
0.5-10μl |
0.1μl |
10 | 1.00% | 0.1 | 0.80% | 0.08 |
5 | 1.50% | 0.075 | 1.50% | 0.075 | |||
1 | 2.50% | 0.025 | 1.50% | 0.015 | |||
MP20A |
2-20μl |
0.5μl |
20 | 0.90% | 0.18 | 0.40% | 0.08 |
10 | 1.20% | 0.12 | 1.00% | 0.1 | |||
2 | 3.00% | 0.06 | 2.00% | 0.04 | |||
MP50A |
5-50μl |
0.5μl |
50 | 0.60% | 0.3 | 0.30% | 0.15 |
25 | 0.90% | 0.225 | 0.60% | 0.15 | |||
5 | 2.00% | 0.1 | 2.00% | 0.1 | |||
MP100A |
10-100μl |
1μl |
100 | 0.80% | 0.8 | 0.15% | 0.15 |
50 | 1.00% | 0.5 | 0.40% | 0.2 | |||
10 | 3.00% | 0.3 | 1.50% | 0.15 | |||
MP20/200A |
20-200μl |
1μl |
200 | 0.60% | 1.2 | 0.15% | 0.3 |
100 | 0.80% | 0.8 | 0.30% | 0.3 | |||
20 | 3.00% | 0.6 | 1.00% | 0.2 | |||
MP50/200A |
50-200μl |
1μl |
200 | 0.60% | 1.2 | 0.15% | 0.3 |
100 | 0.80% | 0.8 | 0.30% | 0.3 | |||
50 | 1.00% | 0.5 | 0.40% | 0.2 | |||
MP100/1000A |
100-1000μl |
5μl |
1000 | 0.60% | 6 | 0.20% | 2 |
500 | 0.70% | 3.5 | 0.25% | 1.25 | |||
100 | 2.00% | 2 | 0.70% | 0.7 | |||
MP200/1000A |
200-1000μl |
5μl |
1000 | 0.60% | 6 | 0.20% | 2 |
500 | 0.70% | 3.5 | 0.25% | 1.25 | |||
200 | 0.90% | 1.8 | 0.30% | 0.6 | |||
MP5000A |
1000-5000μl |
50μl |
5000 | 0.50% | 25 | 0.15% | 7.5 |
2500 | 0.60% | 15 | 0.30% | 7.5 | |||
1000 | 0.70% | 7 | 0.30% | 3 |
8-channel Adjustable Volume Pipettes | ||||||||
Model No | Volume Range | Increment | Test Volume (μl)
|
Maximum Permissible
Systematic error |
Maximum Permissible
Random error |
|||
% | μl | % | μl | |||||
MP10/8A |
0.5-10μl |
0.1μl |
10 | 1.50% | 0.15 | 1.50% | 0.15 | |
5 | 2.50% | 0.125 | 2.50% | 0.125 | ||||
1 | 4.00% | 0.04 | 4.00% | 0.04 | ||||
MP50/8A |
5-50μl |
0.5μl |
50 | 1.00% | 0.5 | 0.50% | 0.25 | |
25 | 1.50% | 0.375 | 1.00% | 0.25 | ||||
5 | 3.00% | 0.15 | 2.00% | 0.1 | ||||
MP300/8A |
50-300μl |
5μl |
300 | 0.70% | 2.1 | 0.25% | 0.75 | |
150 | 1.00% | 1.5 | 0.50% | 0.75 | ||||
50 | 1.50% | 0.75 | 0.80% | 0.4 |
12-channel Adjustable Volume Pipettes | |||||||
Model No | Volume Range | Increment | Test Volume (μl)
|
Maximum Permissible
Systematic error |
Maximum Permissible
Random error |
||
% | μl | % | μl | ||||
MP10/12A |
0.5-10μl |
0.1μl |
10 | 1.50% | 0.15 | 1.50% | 0.15 |
5 | 2.50% | 0.125 | 2.50% | 0.125 | |||
1 | 4.00% | 0.04 | 4.00% | 0.04 | |||
MP50/12A |
5-50μl |
0.5μl |
50 | 1.00% | 0.5 | 0.50% | 0.25 |
25 | 1.50% | 0.375 | 1.00% | 0.25 | |||
5 | 3.00% | 0.15 | 2.00% | 0.1 | |||
MP300/12A |
50-300μl |
5μl |
300 | 0.70% | 2.1 | 0.25% | 0.75 |
150 | 1.00% | 1.5 | 0.50% | 0.75 | |||
50 | 1.50% | 0.75 | 0.80% | 0.4 |
Fixed Volume Pipettes | |||||||
Model No | Volume Range | Increment | Test Volume (μl)
|
Maximum Permissible
Systematic error |
Maximum Permissible
Random error |
||
% | μl | % | μl | ||||
MP5AF | 5μl | – | 5μl | 1.3% | 0.065 | 1.2% | 0.06 |
MP10AF | 10μl | – | 10μl | 0.8% | 0.08 | 0.8% | 0.08 |
MP20AF | 20μl | – | 20μl | 0.6% | 0.12 | 0.5% | 0.1 |
MP25AF | 25μl | – | 25μl | 0.5% | 0.125 | 0.3% | 0.075 |
MP50AF | 50μl | – | 50μl | 0.5% | 0.25 | 0.3% | 0.15 |
MP100AF | 100μl | – | 100μl | 0.5% | 0.5 | 0.3% | 0.3 |
MP200AF | 200μl | – | 200μl | 0.4% | 0.8 | 0.2% | 0.4 |
MP250AF | 250μl | – | 250μl | 0.4% | 1.0 | 0.2% | 0.5 |
MP500AF | 500μl | – | 500μl | 0.3% | 1.5 | 0.2% | 1.0 |
MP1000AF | 1000μl | – | 1000μl | 0.3% | 3.0 | 0.2% | 2.0 |
MP2000AF | 2000μl | – | 2000μl | 0.3% | 6.0 | 0.15% | 3.0 |
MP5000AF | 5000μl | – | 5000μl | 0.3% | 15 | 0.15% | 7.5 |
Note: It is recommended to validate the performance of the pipette after each cycle in the autoclave. It is also recommended to grease the piston and seal of the pipette after 10 cycles in the autoclave.