MODEL SPECIFICATIONS:
| Volume Range | Increment | Test Volume (μl) | Capacity Tolerance (%) | Test Repeatability (%) |
|
0.5-10μl |
0.1μl |
1 | ±2 | ≤3 |
| 5 | ±1.5 | ≤2 | ||
| 10 | ±1 | ≤2 | ||
|
2-20μl |
0.1μl |
2 | ±2 | ≤3 |
| 10 | ±1.5 | ≤2 | ||
| 20 | ±1 | ≤2 | ||
|
10-100μl |
1μl |
10 | ±2 | ≤2 |
| 50 | ±1.5 | ≤1 | ||
| 100 | ±1 | ≤1 | ||
|
20-200μl |
1μl |
20 | ±2 | ≤2 |
| 100 | ±1.5 | ≤1 | ||
| 200 | ±1 | ≤1 | ||
|
100-1000μl |
5μl |
100 | ±2 | ≤1 |
| 500 | ±1 | ≤0.5 | ||
| 1000 | ±1 | ≤0.5 |