Benchtop Dissolved Oxygen Meter

FEATURES: 

  • 1 – 2 points calibration using air-saturated water and zero oxygen solution
  • Salinity and barometric pressure compensation improve the accuracy of the measurement
  • Automatic Temperature Compensation ensures accurate readings
  • Auto read function senses and locks the measurement endpoint
  • Setup menu allows setting the number of calibration points, concentration unit, temperature unit etc
  • Calibration reminder
  • Reset function available to default instrument to factory settings
  • Expanded memory up to 500 readings (only MA-PH980)
  • USB communication interface for data transfer and timed interval readings (only MA-PH980)

Download Product PDF

SKU: N/A Category:

Accessories:              

  • DO100 Dissolved oxygen electrode
  • Electrolyte solution
  • Membrane cap
  • DC9V power adapter

MODEL SPECIFICATIONS:

Model MA-pH810 MA-pH980
Dissolved Oxygen Range 0.0~20.0mg/L (or ppm)
Dissolved Oxygen Accuracy ±0.5mg/L ±0.2mg/L
Dissolved Oxygen Resolution 0.1mg/L 0.01, 0.1mg/L, selectable
% Saturation Range 0.0~200.0%
% Saturation Accuracy ±2.0%
% Saturation Resolution 1% 0.1, 1%, selectable
Temp Range 0~60°C/32~140°F
Temp. Accuracy ±1°C/±1.8°F ±0.5°C/±0.9°F
Temp. Resolution 0.1°C/0.1°F
Temp. Offset Calibration 1 point, reading ±10°C
Dissolved Oxygen Calibration 1 or 2 points
Temp. Compensation 0~40°C, automatic 0~50°C, automatic
Barometric Pressure Correction 60.0 ~ 112.5kPa/450 ~ 850 mmHg, manual
Salinity Correction 0.0 ~ 35.0g/L, manual 0.0 ~ 50.0g/L, manual
Stability Criteria / Low or High
Calibration Due Alarm / 1 to 31 days or Off
Hold Function Manual or auto-endpoint
Memory / 500 data sets
Communication Interface / USB
Connector 6-pin mini-DIN plug
Display Custom LCD (120*60mm) Custom LCD (125*100mm)
Power Requirements DC9Vpower adapter DC5V power adapter
Dimensions (W*D*H) mm 210*205*75 210*188*60
Package Size (W*D*H) mm 310*240*180
Net Weight (kg) 1.5
Gross Weight (kg) 1.7

 

Model

MA-PH810, MA-PH980

Scroll to Top