MODEL SPECIFICATIONS:
| Model | MA-1200 | |
|
Overall Performance |
Throughput | 800 tests/hour, 1200 tests/hour with ISE (Optional) |
| Analysis Method | End-point, Fixed-time, Rate (Kinetic) | |
| Certification | CE, FDA, ISO9001, ISO14001, ISO13485 | |
|
Sample & Reagent System |
Sample Positions | 150 sample positions |
| Reagent Positions | 180 reagent positions | |
| Sample Volume | 1~70μl | |
| Reagent Volume | 10~350μl | |
| Probe | Teflon coating, with anti-collision function, liquid level detection function | |
| Probe Washing | Automatic washing interior and exterior | |
| Reagent Cooling | Refrigerated tray with independent switch | |
|
Reaction System |
Temperature Control | Water bath, 37±0.1℃, real-time monitoring |
| Cuvettes | 160 re-usable cuvettes, optical length 6mm | |
| Mixer probe | Independent stirring | |
| Washing | Automatic cuvettes washing | |
| STAT function | Yes | |
|
Optical System |
Light Source | 12V/20W halogen lamp/fibre bragg grating optional |
| Spectrophotometry | Post-spectral spectrophotometry | |
| Wavelength | 12 wavelengths from 340~800nm | |
| Absorbance | 0~3.5 Abs | |
|
Calibration and QC |
Calibration | Linear: K factor, 1-point , 2-point and multipoint linear
Non-Linear: Spline, Polygon, Index, Ogarithm, Logit-4P, Logit-5P |
| Quality Control | Real-time QC, Westgard multi rule, Cumulative Sum Check, Twin Plot (2D) | |
|
Data Management |
Software | Windows 7/8/10 , 32 or 64 bit |
| LIS System | Bi-direction, support HL7 protocol | |
| Interface | LAN port access | |
| Printer | External, multiple reporting mode available | |
|
Working Conditions |
Power Supply | AC220V±10%, 50/60Hz, 110V±10%, 60Hz, 1500W |
| Temperature | 15~30℃ (±2℃) | |
| Water Consumption | Deionized water: ˂25L/H | |
| Humidity | 40%~85% | |
|
Size & Weight |
External Size (W*D*H) | 1360*795*1155mm |
| Net Weight (kg) | 290 | |
| Package Size (W*D*H) | 1718*958*1427mm | |
| Gross Weight (kg) | 400 | |