MicroPette Plus Autoclavable Pipette

Fully Autoclavable: The MicroPette Plus Pipettes can be fully autoclaved at 121℃, 1 atm for 20 minutes. After autoclaving the pipette must be cooled down and dried over 12 hours.

FEATURES: MICROPETTE PLUS

  • Fully Autoclavable
  • Light weight, ergonomic, low force design
  • Digital display
  • Volume range of 0.1μl to 10ml
  • Easy to calibrate and maintain
  • Design helps avoid repetitive strain injuries
  • Manufactured from innovative materials

FEATURES: MULTI-CHANNEL MICROPETTE PLUS

  • 8 and 12 channel pipettes are available for standard 96-well plates
  • Individual piston and tip cone assemblies allow for easy repair and maintenance
  • Compound material tip cone design allows visual seal verification
  • Universal pipette tips can be used
  • Dispensing head rotates for optimum pipetting convenience

Download Product PDF

 

 

Category:

MODEL SPECIFICATIONS:

Single-channel Adjustable Volume Pipettes
Model No Volume Range Increment Test Volume (μl)

 

Maximum Permissible

Systematic error

Maximum Permissible

Random error

% μl % μl
 

MP2.5A

 

0.1-2.5μl

 

0.05μl

2.5 2.50% 0.0625 2.00% 0.05
1.25 3.00% 0.0375 3.00% 0.0375
0.25 12.00% 0.03 6.00% 0.015
 

MP10A

 

0.5-10μl

 

0.1μl

10 1.00% 0.1 0.80% 0.08
5 1.50% 0.075 1.50% 0.075
1 2.50% 0.025 1.50% 0.015
 

MP20A

 

2-20μl

 

0.5μl

20 0.90% 0.18 0.40% 0.08
10 1.20% 0.12 1.00% 0.1
2 3.00% 0.06 2.00% 0.04
 

MP50A

 

5-50μl

 

0.5μl

50 0.60% 0.3 0.30% 0.15
25 0.90% 0.225 0.60% 0.15
5 2.00% 0.1 2.00% 0.1
 

MP100A

 

10-100μl

 

1μl

100 0.80% 0.8 0.15% 0.15
50 1.00% 0.5 0.40% 0.2
10 3.00% 0.3 1.50% 0.15
 

MP20/200A

 

20-200μl

 

1μl

200 0.60% 1.2 0.15% 0.3
100 0.80% 0.8 0.30% 0.3
20 3.00% 0.6 1.00% 0.2
 

MP50/200A

50-200μl  

1μl

200 0.60% 1.2 0.15% 0.3
100 0.80% 0.8 0.30% 0.3
50 1.00% 0.5 0.40% 0.2
 

MP100/1000A

 

100-1000μl

 

5μl

1000 0.60% 6 0.20% 2
500 0.70% 3.5 0.25% 1.25
100 2.00% 2 0.70% 0.7
 

MP200/1000A

 

200-1000μl

 

5μl

1000 0.60% 6 0.20% 2
500 0.70% 3.5 0.25% 1.25
200 0.90% 1.8 0.30% 0.6
 

MP5000A

 

1000-5000μl

 

50μl

5000 0.50% 25 0.15% 7.5
2500 0.60% 15 0.30% 7.5
1000 0.70% 7 0.30% 3

 

8-channel Adjustable Volume Pipettes
Model No Volume Range Increment Test Volume (μl)

 

Maximum Permissible

Systematic error

Maximum Permissible

Random error

% μl % μl
 

MP10/8A

 

0.5-10μl

 

0.1μl

10 1.50% 0.15 1.50% 0.15
5 2.50% 0.125 2.50% 0.125
1 4.00% 0.04 4.00% 0.04
 

MP50/8A

 

5-50μl

 

0.5μl

50 1.00% 0.5 0.50% 0.25
25 1.50% 0.375 1.00% 0.25
5 3.00% 0.15 2.00% 0.1
 

MP300/8A

50-300μl  

5μl

300 0.70% 2.1 0.25% 0.75
150 1.00% 1.5 0.50% 0.75
50 1.50% 0.75 0.80% 0.4

 

12-channel Adjustable Volume Pipettes
Model No Volume Range Increment Test Volume (μl)

 

Maximum Permissible

Systematic error

Maximum Permissible

Random error

% μl % μl
 

MP10/12A

 

0.5-10μl

 

0.1μl

10 1.50% 0.15 1.50% 0.15
5 2.50% 0.125 2.50% 0.125
1 4.00% 0.04 4.00% 0.04
 

MP50/12A

 

5-50μl

 

0.5μl

50 1.00% 0.5 0.50% 0.25
25 1.50% 0.375 1.00% 0.25
5 3.00% 0.15 2.00% 0.1
 

MP300/12A

 

50-300μl

 

5μl

300 0.70% 2.1 0.25% 0.75
150 1.00% 1.5 0.50% 0.75
50 1.50% 0.75 0.80% 0.4

 

Fixed Volume Pipettes
Model No Volume Range Increment Test Volume (μl)

 

Maximum Permissible

Systematic error

Maximum Permissible

Random error

% μl % μl
MP5AF 5μl 5μl 1.3% 0.065 1.2% 0.06
MP10AF 10μl 10μl 0.8% 0.08 0.8% 0.08
MP20AF 20μl 20μl 0.6% 0.12 0.5% 0.1
MP25AF 25μl 25μl 0.5% 0.125 0.3% 0.075
MP50AF 50μl 50μl 0.5% 0.25 0.3% 0.15
MP100AF 100μl 100μl 0.5% 0.5 0.3% 0.3
MP200AF 200μl 200μl 0.4% 0.8 0.2% 0.4
MP250AF 250μl 250μl 0.4% 1.0 0.2% 0.5
MP500AF 500μl 500μl 0.3% 1.5 0.2% 1.0
MP1000AF 1000μl 1000μl 0.3% 3.0 0.2% 2.0
MP2000AF 2000μl 2000μl 0.3% 6.0 0.15% 3.0
MP5000AF 5000μl 5000μl 0.3% 15 0.15% 7.5

 

Note: It is recommended to validate the performance of the pipette after each cycle in the autoclave. It is also recommended to grease the piston and seal of the pipette after 10 cycles in the autoclave.

Scroll to Top