MODEL SPECIFICATIONS:
Model | MA-III 15 | MA-III 30 | |
Display | LCD | ||
Water Output Type | RO and DI Water | ||
Water Output Speed | 15 L/H | 30 L/H | |
Purifying Procedure | PF+AC+RO+DI | ||
Water Supply Requirement | Tap water: TDS<200ppm, 5~45°C, 1.0~4.0Kgf/cm2
(If inlet TDS>200ppm, extra pre-treatment filter is recommended) |
||
Purification System |
Pre-Treatment Unit | PP filter*1; Special active carbon filter*2 | 5μm PP deep filter, Special active carbon filter*2 |
RO Unit | 100GPD RO membrane*1 | 200GPD RO membrane*1 | |
Subsequent Unit | Deionisation purification column (imported resin)*2 | Deionisation purification column (imported resin)*3 | |
Pure Water Quality |
Ion Rejection Rate | 97%~99% (new RO membrane) | |
Organic Rejection Rate | >99%, when MW>200 Dalton | ||
Particles and Bacteria Rejection Rate | >99% | ||
Resistivity | 15~18.2MΩ.cm | ||
Conductivity | 0.055~0.067μs/cm | ||
Control System | Automatic electronic pressure sensor, RO membrane automatically flushes, Automatic stop when
shortage of water supply occurs, Automatic stop when water tank is full, Automatically cuts off when pump stops, 24-hour operation |
||
Flow Rate | 2.0 L/min (when there is water in the water tank) | ||
Water Quality Monitor | TDS/conductivity test pen | ||
Consumption | 72W | ||
Power Supply | AC220V±10%, 50Hz+5% (standard); AC110V, 60Hz (optional) | ||
Standard Configuration | Main body (includes 1 set of purification column), 15L tank, TDS /conductivity test pen, accessory pack | ||
External Size (W*D*H) | 410*220*420mm | ||
Package Size (W*D*H) | 600*490*580mm | ||
Gross Weight (kg) | 35 |